* LMIS empty
 
* LMIS empty
 
** Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder
 
** Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder
 +
* Sources were refilled?
 
* Detector tube bad (outgassed)
 
* Detector tube bad (outgassed)
 
** Replaced
 
** Replaced
Line 31: Line 32:
 
* Faulty power supply
 
* Faulty power supply
 
** Repaired module (replaced resistor?)
 
** Repaired module (replaced resistor?)
 
+
* Bad seal on pneumatic stabalizer
 +
** Bought (silicone?) sheet to repair
 +
* Turbo pump sometimes noisy, but seems to still work
  
 
== GIS ==
 
== GIS ==

Revision as of 00:10, 13 February 2019

This seems to be a pretty popular FIB. Based on the XL30?

Refilling

Videos for GIS refill: https://www.youtube.com/channel/UCebcTTg_Al48R9Uxdy8sS9Q

Chemicals Markus is using to refill (TODO: get image of Ga bottle, process):

Markus fei fib200 chem merk hexacarbonyl-w.jpg Markus fei fib200 chem sa tetraethyl orthosilicate.jpg


Calibration

General procedure is to wobble the focus and minimize how much drift occurs when sweeping. Looks something like this: https://youtu.be/Ln2GDMw-Vf8


Markus

mcmaster: took a bunch of notes, but they were lost. I still have the pictures though

Problems when acquired:

  • LMIS empty
    • Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder
  • Sources were refilled?
  • Detector tube bad (outgassed)
    • Replaced
  • XeF2 leak. Damaged stage gears
    • Removed XeF2 GIS and blanked off
    • Bought similar replacement gears, machined to fit
    • TOOD: install new seals. Requires special XeF2 rated version not in other GIS
  • Faulty power supply
    • Repaired module (replaced resistor?)
  • Bad seal on pneumatic stabalizer
    • Bought (silicone?) sheet to repair
  • Turbo pump sometimes noisy, but seems to still work

GIS

GIS cluster. Note one of the injectors missing (more on it below)

Markus fei fib200 gis cluster1.jpg Markus fei fib200 gis cluster2.jpg Markus fei fib200 gis cluster3.jpg

A new Pt injector tip:

Markus fei fib200 gis pt-new btm.jpg Markus fei fib200 gis pt-new side.jpg Markus fei fib200 gis pt-new side2.jpg Markus fei fib200 gis pt-new top.jpg

This XeF2 injector sat for a long time in storage. The XeF2 ate through the seals and caused considerable damage in the FIB

Markus fei fib200 gis xef2-bad 1.jpg Markus fei fib200 gis xef2-bad 2.jpg Markus fei fib200 gis xef2-bad 3.jpg


Computer

Markus fei fib200 computer console.jpg

Markus fei fib200 computer front1.jpg Markus fei fib200 computer front2.jpg Markus fei fib200 computer front3.jpg

Markus fei fib200 computer back.jpg


Misc

Aperture motor test: https://youtu.be/TIqlTjR2UH4

Test edit after calibration (cut trace):

Markus fei fib200 edit1.jpg Markus fei fib200 edit2.jpg