Line 23: | Line 23: | ||
* LMIS empty | * LMIS empty | ||
** Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder | ** Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder | ||
+ | * Sources were refilled? | ||
* Detector tube bad (outgassed) | * Detector tube bad (outgassed) | ||
** Replaced | ** Replaced | ||
Line 31: | Line 32: | ||
* Faulty power supply | * Faulty power supply | ||
** Repaired module (replaced resistor?) | ** Repaired module (replaced resistor?) | ||
− | + | * Bad seal on pneumatic stabalizer | |
+ | ** Bought (silicone?) sheet to repair | ||
+ | * Turbo pump sometimes noisy, but seems to still work | ||
== GIS == | == GIS == |
Revision as of 00:10, 13 February 2019
This seems to be a pretty popular FIB. Based on the XL30?
Refilling
Videos for GIS refill: https://www.youtube.com/channel/UCebcTTg_Al48R9Uxdy8sS9Q
Chemicals Markus is using to refill (TODO: get image of Ga bottle, process):
Calibration
General procedure is to wobble the focus and minimize how much drift occurs when sweeping. Looks something like this: https://youtu.be/Ln2GDMw-Vf8
Markus
mcmaster: took a bunch of notes, but they were lost. I still have the pictures though
Problems when acquired:
- LMIS empty
- Fabricated replacement using Ga from Sigma-Aldrich, tungsten wire, syringe, and DIY spot welder
- Sources were refilled?
- Detector tube bad (outgassed)
- Replaced
- XeF2 leak. Damaged stage gears
- Removed XeF2 GIS and blanked off
- Bought similar replacement gears, machined to fit
- TOOD: install new seals. Requires special XeF2 rated version not in other GIS
- Faulty power supply
- Repaired module (replaced resistor?)
- Bad seal on pneumatic stabalizer
- Bought (silicone?) sheet to repair
- Turbo pump sometimes noisy, but seems to still work