Difference between revisions of "Hamamatsu"
(Tag: Visual edit) |
(Tag: Visual edit) |
||
Line 22: | Line 22: | ||
|C790-10 | |C790-10 | ||
| | | | ||
+ | |X-ray | ||
+ | |Unmarked | ||
| | | | ||
− | |||
| | | | ||
| | | | ||
| | | | ||
− | |||
|- | |- | ||
|C2400-03 | |C2400-03 | ||
Line 61: | Line 61: | ||
| | | | ||
|Advertised in new catalog as the only modern general vidicon supported | |Advertised in new catalog as the only modern general vidicon supported | ||
+ | |- | ||
+ | |C2400-08 | ||
+ | | | ||
+ | | | ||
+ | | | ||
+ | | | ||
+ | | | ||
+ | | | ||
+ | |McMaster | ||
+ | |||
+ | |||
+ | Tube hard to get to, but could take out if really determined. Would need to remove either (fiber bundle connector?) or most of the electronics | ||
|- | |- | ||
|C2400-53 | |C2400-53 | ||
Line 83: | Line 95: | ||
|- | |- | ||
|C2741- | |C2741- | ||
− | |||
| | | | ||
| | | | ||
Line 95: | Line 106: | ||
| | | | ||
| | | | ||
− | | | + | |McMaster weird poorly marked unit |
Line 125: | Line 136: | ||
| | | | ||
|[http://alacron.com/clientuploads/directory/Cameras/HAMAMATSU/c2741-03-Datasheet.pdf Datasheet] | |[http://alacron.com/clientuploads/directory/Cameras/HAMAMATSU/c2741-03-Datasheet.pdf Datasheet] | ||
+ | |||
+ | |||
+ | McMaster has a few | ||
|- | |- | ||
|C2741-03B | |C2741-03B | ||
Line 245: | Line 259: | ||
<br /> | <br /> | ||
− | == McMaster == | + | ==McMaster== |
− | === 2019-12-09 === | + | ===2019-12-09=== |
Summary | Summary | ||
Revision as of 08:13, 11 December 2019
This page was created to understand some Vidicon repair and interchanagability
C2400-16 1-inch Super- Harpicon. 400 to 700 nm, 0.001 to 10 lux
Vidicon
Heads:
Model | Connector | For | Tube | Resolution | Sensitivity | Yoke | Notes |
---|---|---|---|---|---|---|---|
C790-10 | X-ray | Unmarked | |||||
C2400-03 | IR
800-2100 nm 400-1800 nm |
1" | 720 x 576
600 line |
0.1-10 lux | KV-5767 | Lens: Xenoplan 1,4/23 mm
Advertised in new catalog as the only modern IR vidicon supported | |
C2400-05 | UV
250-750 nm |
1" | 700 line | 0.01-1 lux | |||
C2400-07 | Visible microscope | Newvicon / Matsushita | Advertised in new catalog as the only modern general vidicon supported | ||||
C2400-08 | McMaster
| ||||||
C2400-53 | IR
400-1800 |
2/3" | 550 line | 0.1-10 lux | |||
C2400-55 | UV
250-750 |
2/3" | 700 line | 0.01-1 lux | |||
C2741- | 1"
NEWVICON S4076 MATSUSHITA |
McMaster weird poorly marked unit
| |||||
C2741-03 | IR
400 to 1800 nm 400 to 2200 nm |
N2606
N2606-06 1"
|
600 horizontal
CCIR: 625x509 |
S/N: 46 dB p-p/rms | Datasheet
| ||
C2741-03B | 1800 nm | N2606-02 | 22/01/2008
£8,340 Package includes Camera Head, Controller Unit, Camera Cable & BNC to BNC Cable
PC2Vision Frame Grabber Card & Cable Set Wasabi for Video Acquisition & Analysis Software |
C2400 and C2741 seem generally compatible. The one C2400 camera I have has a
https://www.artisantg.com/info/PDF__48616D616D617473755F43323430305F43616D6572615F446174617368656574.pdf
C790 X-Ray camera
moved from https://nucwiki.org/wiki/index.php/Hamamatsu_C790
mcmaster: got "X-RAY TV CAMERA" off of eBay missing cable. Both the camera and controller are marked C790-10 Was able to source something close to the needed CMC connectors and was able to retool for the proper cable. Now need to wire between them so I can boot the unit
Some relationship to Softex. Maybe they are used in Softex machines?
All wires on connector are used (at least on camera side).
Japan imports DDK JMSP2528 connector 025-028 silver-plated 28 core aviation plug 4 thick needle 24 thin needle
"Hamamatsu Softex X-Ray TV Camera controller C3205": seems to use the same connector
"Hamamatsu C6086 X-Ray Camera Controller": for CCD I think, not compatible
Head: MODEL C790, TYPE 10, SER.NO 88054284
Controller: MODEL C790, TYPE 10, SER.NO 88074380
Camera
Appears to be vidicon? Hard to see...in a black sub-enclosure. Enclosure is marked
YOKE ASSEMBLY XVC-492E HC-01 CHUOMUSEN CO., LTD. MADE IN JAPAN 378024
Boards:
- OXVPOI-2 PRE AMP
- OXPPOI-2 HV-POWER
- OXVPO2-2 DEF FOCUS
Maybe 22 or 24 AWG on connector. Power pins don't appear to use a larger gauge than the rest.
CCD
Model | Connector | For | Tube | Resolution | Yoke | Notes |
---|---|---|---|---|---|---|
C5840-10: EIA
C5840-11: CCIR |
IR
400 to 1650 nm |
2/3"? | Datasheet
Datasheet has graph comparing to C2741-03. TLDR: this is much less sensitive |
C2741-63 camera controller
- 12PinCCD camera connector
- RS-232C
- 12PIN cable 5 m A3194-01
C6086 Series X-Ray CCD Camera
- 25 μm (C6086-03, 04)
- 30 μm (C6086-53, 54)
- 43 μm (C6086-13, 14)
- 62 μm (C6086-23, 24)
- Effective X-ray tube voltage range
- 10 to 100 kVp
https://www.artisantg.com/info/PDF__48616D616D617473755F43323430305F33325F33355F446174617368656574.pdf
https://www.artisantg.com/info/ATGckodl.pdf
C2400-77 2/3-inch Interline CCD
C2400-75 1/2-inch Interline CCD
McMaster
2019-12-09
Summary
-Composite capture working okay
-Pre-amp currently in iffy causes flicker
originally in "bad"
-Both C2741-03 tubes are slow response
-Each C2741-03 tube has a defect as currently used
Esj 54: poor SNR
Esj 31: uneven illumination
can be trimmed?
I believe they exhibit the same defect no matter the unit put into
-All 3 units use KV-5767 yoke assembly
-Pre-amps look like they are interchangable between C2741-03 and C2741-, but appear not to be
-Different tubes
C2741-03: N2606-10
C2741-: S4076
-Haven't touched C2400
-Haven't compard to
https://www.ebay.com/itm/4-Used-Tubes-3-Newvicon-S4076-Vidicon-Tube-One-RCA-8507A-Untested-But-Guaranteed/383310410334?hash=item593f14b25e:g:ByAAAOSwCL9d6HFD
4 Used Tubes-3 Newvicon S4076 Vidicon Tube One RCA 8507A Untested But Guaranteed
probably should get these to play with