Difference between revisions of "Molectron"
(Tag: Visual edit) |
(Tag: Visual edit) |
||
(2 intermediate revisions by the same user not shown) | |||
Line 7: | Line 7: | ||
!Connector | !Connector | ||
!Sensor | !Sensor | ||
− | ! | + | !Compatible |
+ | Meters | ||
!Memory | !Memory | ||
! | ! | ||
|- | |- | ||
− | | | + | |J |
|DB25 | |DB25 | ||
− | |J8 | + | |Joule: |
+ | J8 | ||
J25 | J25 | ||
J50 | J50 | ||
+ | |||
+ | J100 | ||
+ | |||
+ | |||
+ | Power: | ||
OP-2 | OP-2 | ||
Line 29: | Line 36: | ||
|25LC320 | |25LC320 | ||
(SPI) | (SPI) | ||
− | |Joule sensor: J25 is the most common | + | |Joule sensor: J25 is the most common followed by J50. J8 seems rare |
+ | |||
+ | Power sensor: OP-2 might be the most common. S10 seems rare | ||
− | + | I've been calling these JSA due to "JSA-BNC", but that name might not be correct | |
|- | |- | ||
|LM | |LM | ||
+ | (LabMax) | ||
|DB25 | |DB25 | ||
|LM-2 | |LM-2 | ||
Line 44: | Line 54: | ||
LM-100 | LM-100 | ||
− | | | + | |4omega?<br /> |
|24C04 | |24C04 | ||
(I2C) | (I2C) | ||
| | | | ||
+ | |- | ||
+ | |LM-P | ||
+ | |DB25 | ||
+ | |LM-P10 | ||
+ | | | ||
+ | | | ||
+ | |LM-P EEPROM looks slightly different than LM and in fact the LabMax-TOP user manual mentions these sensors are not compatible | ||
|- | |- | ||
|PM | |PM | ||
+ | (PowerMax) | ||
|DB25 | |DB25 | ||
|PM10 | |PM10 | ||
PM150 | PM150 | ||
+ | |||
+ | PM-P (PowerMax-Pro) | ||
|PM5200 | |PM5200 | ||
|93C46N | |93C46N | ||
(SPI) | (SPI) | ||
− | | | + | |Even if it was electrically compatible w/ JSA EEPROM format is completely different |
− | Even if it was electrically compatible w/ JSA EEPROM format is completely different | ||
|} | |} | ||
Line 90: | Line 109: | ||
*FieldMax-TOP | *FieldMax-TOP | ||
*Fieldmaster | *Fieldmaster | ||
+ | *LabMax-TOP | ||
+ | **Compatible with LM-, PM-, and PS- model thermopile sensors, and OP-2 and LM-2 model optical sensors. No adapters required. | ||
+ | **Measures energy to 10000 pps and captures every pulse to 1000 pps with J-model and EnergyMax sensors | ||
+ | **LabMax is not compatible with the legacy LM-P model energy sensors <br /> | ||
+ | |||
+ | PM official pinout for thermopile sensors | ||
+ | |||
+ | * COHERENT_LaserPowerSensors_FlowRequirements_FittingSpecs_PinAssignments_04-2017mp.pdf | ||
+ | * https://laserpointerforums.com/threads/help-with-lpm-wiring-db25-pinout-pm5200-to-attach-thermopile-sensor.107079/page-2 | ||
+ | |||
Lots of good info: | Lots of good info: |
Latest revision as of 05:11, 12 November 2021
Information on Molectron / coherent laser power meters. Ultimately for the ezlaze project. I'm mostly interested in pyroelectric heads for pulsed power.
Contents
Ecosystem
Family | Connector | Sensor | Compatible
Meters |
Memory | |
---|---|---|---|---|---|
J | DB25 | Joule:
J8 J25 J50 J100
OP-2 S10 |
3sigma
FieldMax-TOP FieldMax-TO FieldMate |
25LC320
(SPI) |
Joule sensor: J25 is the most common followed by J50. J8 seems rare
Power sensor: OP-2 might be the most common. S10 seems rare
|
LM
(LabMax) |
DB25 | LM-2
LM-10 LM-30 LM-45 LM-100 |
4omega? |
24C04
(I2C) |
|
LM-P | DB25 | LM-P10 | LM-P EEPROM looks slightly different than LM and in fact the LabMax-TOP user manual mentions these sensors are not compatible | ||
PM
(PowerMax) |
DB25 | PM10
PM150 PM-P (PowerMax-Pro) |
PM5200 | 93C46N
(SPI) |
Even if it was electrically compatible w/ JSA EEPROM format is completely different |
Ecosystems:
- J DB-25 smart probes
- Interface: DB25
- Also known as JSA?
- 3sigma, FieldMax-TOP, FieldMax-TO, FieldMate
- Intended to be used with J25, J50 power probes
- I also have an S10 energy probe, seems to be compatible
- LM DB-25 smart probes
- Interface: DB25
- Intended to be used with LM probes
Warning: J25 and LM are not compatible.
Other DB25 probes (none are known to be compatible):
- Ophir
- Laserpoint
- Gentec
- Newport
List of Molectron related power meters:
- 3sigma
- 4omega
- FieldMate
- FieldMax-TO
- FieldMax-TOP
- Fieldmaster
- LabMax-TOP
- Compatible with LM-, PM-, and PS- model thermopile sensors, and OP-2 and LM-2 model optical sensors. No adapters required.
- Measures energy to 10000 pps and captures every pulse to 1000 pps with J-model and EnergyMax sensors
- LabMax is not compatible with the legacy LM-P model energy sensors
PM official pinout for thermopile sensors
- COHERENT_LaserPowerSensors_FlowRequirements_FittingSpecs_PinAssignments_04-2017mp.pdf
- https://laserpointerforums.com/threads/help-with-lpm-wiring-db25-pinout-pm5200-to-attach-thermopile-sensor.107079/page-2
Lots of good info:
- Energy Sensors Coherent J25.pdf
- 2009_LMCCatalog86.pdf
There was a conversion box you could buy at one point though (thermal-smartsensor-adapter-data-sheet.pdf), so it might just be different EPROM format
- JSA-BNC Adapter
- 0012-2340: JSA-BNC Adapter for use with 3sigma, FieldMax-TOP, and FieldMax-P
- 1056827: SmartSensor Adapter for use with 3sigma, FieldMax-TOP, FieldMax-TO,and FieldMate
- Automatically configures itself to work with LM-model sensors
J25
TODO: figure out pinout, EEPROM format so I can recal and upgrade adapters
LM
3Sigma
File archive:
3sigmaAppQuickRef.pdf 3sigmaLVDriverUG.pdf 3sigma_LV_Driver.zip 3sigma Users Manual.pdf 3sigmaWinApps.zip
Battery
Ni-MH BATTERY 7.2V 3.8 A-h MOLECTRON P/N 0012-0100
JD1000
Power supply: CS-1203000 (12V @ 3A, center positive)
McMaster
J25LP-4 sensor
Came with 3 sigma in the trash
3Sigma output:
Probe Resp: 1.827E+01 V/J @ 248 nm Probe Cal Date Dec 8 2008
Head label:
Responsivity 1.828e+1 V/J Cal: 248 nm
COHERENT Model: J25LP-4
COHERENT CALIBRATION TAG DATE CALIB 12/08/08 TECH KH NEXT DUE 12/08/09
Item #: 0010-8145 Serial #: 0437E03
Shell label:
COHERENT PROBE SERIAL NO. 0437E03
J25LP sensor
Came with JD1000. BNC output but no 25 pin adapter
Back:
MODEL NO: J25LP SERIAL NO: 238 RESPONSITIVITY 5.10 VOLT/JOULE 337.1 nm
S10 sensor
From 3Sigmal (eBay)
3Sigma output:
Probe Resp: 2.570E-1 A/W @ 514 nm Probe Cal Date Sep 10 2003
DB25 label:
Molectron MODEL NO.: S10 SERIAL NO. 0338H03
DB25 label:
Molectron SER NO 0338H03 CAL DATE 9/18/03 CAL DUE 9/18/04 TECH B DUNLAP 0011-6322 TRACETABLE TO NIST
Cable label: 0012-2530
25LC320
- 32K SPI Bus Serial EEPROM
Pin | Function | Notes |
---|---|---|
1 | Silkscreen: Th | |
2 | Silkscreen: Py | |
3 | Sensor + | Silkscreen: SG |
4 | Silkscreen: S1
W3 to W4 jumper | |
5 | EEPROM CSn | Pullup resistor to VCC |
6 | 4, 6, 14,16 shorted | |
7 | ||
8 | ||
9 | Jumper | N/C |
10 | Jumper | 10-12 shorted |
11 | Jumper | 10-12 shorted |
12 | Jumper | 10-12 shorted |
13 | EEPROM VSS | |
14 | 4, 6, 14,16 shorted | |
15 | Sensor - | |
16 | 4, 6, 14,16 shorted | |
17 | ||
18 | ||
19 | ||
20 | ||
21 | ||
22 | EEPROM SCK | |
23 | EEPROM SI | |
24 | EEPROM SO | |
25 | EEPROM VCC | 2.5-5.5V |
EEPROM pin | Function | DB25 pin | Note |
---|---|---|---|
1 | CSn | 5 | |
2 | SO | 24 | |
3 | WPn | 25 | VCC |
4 | VSS | 13 | |
5 | SI | 23 | |
6 | SCK | 22 | |
7 | HOLDn | 25 | VCC |
8 | VCC | 25 |
J-25MB-HE sensor
Cable label:
COHERENT Item# 1110748 Model: J-25MB-HE Serial#: 0984E07R
Responsitivity 6.694e+0 V/J Cal 1064 nm
LM-10
LM-10 HTD 250-10600nm 10mW-10w 500mJ/cm2 10ns @ 1064n Laser
Relatively cheap impulse buy to compare EEPROM formats
Significantly more complicated inside. Instead of being a simple BNC out there is a whole bundle of cases. Think this is "heated" version maybe though and there might be a dumber version
Atmel 24C04
- 4KB I2C
- PDIP8
- I was going to make an adapter and figure out pinout, but I decided to just clip on instead
Pin | Function | Notes |
---|---|---|
1 | ||
2 | ||
3 | ||
4 | ||
5 | ||
6 | ||
7 | ||
8 | ||
9 | ||
10 | ||
11 | ||
12 | ||
13 | ||
14 | ||
15 | ||
16 | ||
17 | ||
18 | ||
19 | ||
20 | ||
21 | ||
22 | ||
23 | ||
24 | ||
25 |
3Sigma (eBay) meter
Came with CW sensor
3Sigma (dumpster dive) meter
Came w/ sensor + BNC adapter
Screen is messed up
Came with Lemo cable. Doesn't fit trigger ports or anything I can tell
JD1000 meter
Capacitor exploded when plugged in (presumably due to age). Didn't work after, although some lights turned on. Scrapped
Newport 835
Newport project: http://hololaser.kwaoo.me/electronics/Newport835.html
I just happen to also have a few Newport 835s. Unfortunately my inventory is:
- Newport 835, S/N 1477: calibrated for 818-SL, S/N 3190
- Newport 835, S/N 1502: calibrated for 818-SL, S/N 3165
- Newport 818-SL, S/N 3191
...which is close but doesn't entirely match. Dumped the EPROM / calibration data though in case I want to play with it in the future
Liconix 45PM
My oldest laser power meter. Meter + head both have S/N 9110014.
Completely analog / calibrated via pots.
More info
Source code: https://github.com/JohnDMcMaster/molectron