Difference between revisions of "Molectron"

From Microwiki
Jump to: navigation, search
(3Sigma (eBay) sensor)
Line 20: Line 20:
 
Lots of good info:
 
Lots of good info:
  
* Energy Sensors Coherent J25.pdf
+
*Energy Sensors Coherent J25.pdf
* 2009_LMCCatalog86.pdf
+
*2009_LMCCatalog86.pdf
  
  
 
There was a conversion box you could buy at one point though (thermal-smartsensor-adapter-data-sheet.pdf), so it might just be different EPROM format
 
There was a conversion box you could buy at one point though (thermal-smartsensor-adapter-data-sheet.pdf), so it might just be different EPROM format
  
* JSA-BNC Adapter
+
*JSA-BNC Adapter
* 0012-2340: JSA-BNC Adapter for use with 3sigma, FieldMax-TOP, and FieldMax-P
+
*0012-2340: JSA-BNC Adapter for use with 3sigma, FieldMax-TOP, and FieldMax-P
* 1056827: SmartSensor Adapter for use with 3sigma, FieldMax-TOP, FieldMax-TO,and FieldMate
+
*1056827: SmartSensor Adapter for use with 3sigma, FieldMax-TOP, FieldMax-TO,and FieldMate
* Automatically configures itself to work with LM-model sensors
+
*Automatically configures itself to work with LM-model sensors
  
 
<br />
 
<br />
Line 65: Line 65:
 
  337.1 nm
 
  337.1 nm
  
===3Sigma (eBay) sensor===
+
===S10  sensor===
 +
 
 +
From 3Sigmal (eBay)
 +
 
  
 
3Sigma output:
 
3Sigma output:
Line 75: Line 78:
  
 
DB25 label:
 
DB25 label:
  FIXME
+
  MODEL NO.: S10
 
 
Sensor label:
 
FIXME
 
  
 
===J-25MB-HE sensor===
 
===J-25MB-HE sensor===

Revision as of 00:50, 27 October 2021

Information on Molectron / coherent laser power meters. Ultimately for the ezlaze project. I'm mostly interested in pyroelectric heads for pulsed power.

Ecosystem

Ecosystems:

  • J DB-25 smart probes
    • Interface: DB25
    • Also known as JSA?
    • 3sigma, FieldMax-TOP, FieldMax-TO, FieldMate.
    • Intended to be used with J25, J50 power probes
    • I also have an S10 energy probe, seems to be compatible
  • LM DB-25 smart probes
    • Interface: DB25
    • Intended to be used with LM probes

Warning: J25 and LM are not compatible.


Lots of good info:

  • Energy Sensors Coherent J25.pdf
  • 2009_LMCCatalog86.pdf


There was a conversion box you could buy at one point though (thermal-smartsensor-adapter-data-sheet.pdf), so it might just be different EPROM format

  • JSA-BNC Adapter
  • 0012-2340: JSA-BNC Adapter for use with 3sigma, FieldMax-TOP, and FieldMax-P
  • 1056827: SmartSensor Adapter for use with 3sigma, FieldMax-TOP, FieldMax-TO,and FieldMate
  • Automatically configures itself to work with LM-model sensors


J25

TODO: figure out pinout, EEPROM format so I can recal and upgrade adapters

LM

3Sigma

File archive:

3sigmaAppQuickRef.pdf
3sigmaLVDriverUG.pdf
3sigma_LV_Driver.zip
3sigma Users Manual.pdf
3sigmaWinApps.zip


JD1000

Power supply: CS-1203000 (12V @ 3A, center positive)

McMaster

J25LP sensor

Came with JD1000. BNC output but no 25 pin adapter

Back:

MODEL NO: J25LP
SERIAL NO: 238

RESPONSITIVITY
5.10 VOLT/JOULE
337.1 nm

S10 sensor

From 3Sigmal (eBay)


3Sigma output:

Probe Resp: 2.570E-1 A/W @ 514 nm
Probe Cal Date Sep 10 2003
Wavelenth Corr 1064 nm / On
Attenuation Corr: 1.00E0 / Off
Area Corr 1.0000 sq cm / Off

DB25 label:

MODEL NO.: S10

J-25MB-HE sensor

3Sigma output:

FIXME

Cable label:

COHERENT
Item# 1110748
Model: J-25MB-HE
Serial#: 0984E07R
Responsitivity
6.694e + 0 V/J
Cal 1064 nm

LM-10

LM-10 HTD 250-10600nm 10mW-10w 500mJ/cm2 10ns @ 1064n Laser

Relatively cheap impulse buy to compare EEPROM formats

3Sigma (eBay) meter

Came with CW sensor

3Sigma (dumpster dive) meter

Came w/ sensor + BNC adapter

Screen is messed up

Came with Lemo cable. Doesn't fit trigger ports or anything I can tell

JD1000 meter

Capacitor exploded when plugged in (presumably due to age). Didn't work after, although some lights turned on. Scrapped