SEM EDS

From Microwiki
Revision as of 00:20, 25 December 2018 by Test (talk | contribs) (Created page with "Category:SEM "Energy-dispersive X-ray spectroscopy (EDS, EDX, or XEDS)" List of EDS manufacturers: ''' EDAX '''* http:''www.edax.com/|Websit...")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to: navigation, search


"Energy-dispersive X-ray spectroscopy (EDS, EDX, or XEDS)"

List of EDS manufacturers:

EDAX * Website Oxford Instruments * Website Noran Instruments (now Thermo Scientific) Princeton Gamma-Tech (PGT, now Thermo Scientific) Bruker * Website Evex (Kevex?) * "To enjoy this site you'll need to update your Flash Player. It's easy, painless and will take just a moment… " * haha never buying your products Link Analytical e2v Sirius

Typical systme:

Thin window (Be, C, B) SiLi crystal FET pre-amplifier Vacuum Cold finger Pre-amplifier Amplifier MDA electronics (“multi channel analyzer”)

Detector

Styles

Si-Li Silicon Drift Detector (SDD)

Si-Li

Operates at LN2 temps. If operated at room temp it will be destroyed

SDD

"SDD detectors operate at a few tens of degrees below zero"

Window

Styles [epmalab]:

Be foil Supported window Windowless

Be foil

"The most common EDS detector window has been made of Be foil ~7.6 um (0.3 mil) thick. It allows good transmission of X-rays above ~ 1 keV. It is strong enough to withstand venting to atmospheric pressure, and opaque to optical photons." [epmalab]

Supported window

"For transmission of light element X-rays (<1 keV), windows ~0.25 um thick of BN, SiN, diamond or polymer are used. They must use supporting grids to withstand pressure differentials; the grid (e.g., Si or Ni) takes up about 15% of the area, but thin enough that low energy X-rays pass through." [epmalab]

Ultra Thin Window (UTW)

made of:

Polymide ref "poly-vinyl formal (FORMVAR), butvar, parylene, kevlar, polypropylene, lexan or polyimide" ref

Windowless

"no film, and there is a turret that allows swapping with a Be window. Difficult to use as oil or ice can coat the detector surface" [epmalab]

Gamma Spectacular

http:beeresearch.com.au/

How hard would it be to use this as EDS controller?

Model comparison

Image Vendor Model Year Features List price Ask price Manual Schem SW
EDAX Genesis Spectrum 2006 [[http:siliconpr0n.org/media/edax/genesis/GenEDS.pdf Yes]] [[http:siliconpr0n.org/media/edax/genesis/ Yes]]


References

http:www.charfac.umn.edu/instruments/eds_on_sem_primer.pdf * Has a nice CAD drawing explaining the components of real EDS detectors http:epmalab.uoregon.edu/pdfs/Electron Beam MicroAnalysis-Geol_619-8.ppt

\\