Difference between revisions of "Allied MultiPrep"
(Tag: Visual edit) |
(→McMaster spare repair) (Tag: Visual edit) |
||
Line 43: | Line 43: | ||
What is max voltage allied puts out on motors? | What is max voltage allied puts out on motors? | ||
− | * 1-5 | + | *1-5 |
− | ** 7.6 (fixed) | + | **7.6 (fixed) |
− | ** osc | + | **osc |
− | * 2-6 | + | *2-6 |
− | ** 15.0 (max) | + | **15.0 (max) |
− | ** drive | + | **drive |
<br /> | <br /> | ||
Line 227: | Line 227: | ||
| | | | ||
|}<br /> | |}<br /> | ||
+ | <br /> | ||
+ | |||
+ | ==McMaster== | ||
+ | |||
+ | |||
+ | 2020-01-14 | ||
+ | |||
+ | * Win cross sectioning kit | ||
+ | |||
+ | |||
+ | 2020-02-20 | ||
+ | |||
+ | * Order sample holder | ||
+ | * Order replacement tubing | ||
+ | ** Later found I had some spare, but meh can use for other projects | ||
+ | |||
+ | <br /> | ||
+ | |||
+ | ===Primary=== | ||
+ | 2019-12-19 | ||
+ | |||
+ | * Submitted offer. Negotiations failed | ||
+ | |||
+ | |||
+ | 2020-02-19 | ||
+ | |||
+ | * Seller accepted offer (was firm at old offer). Local pick up | ||
+ | |||
+ | |||
+ | 2020-02-232 | ||
+ | |||
+ | * Swap DRO from spare to primary machine | ||
+ | |||
− | |||
2020-02-23 | 2020-02-23 | ||
− | * Observe osc motor is not moving | + | *Observe osc motor is not moving |
− | * Observe osc doesn't have slit like spare to keep in place...how is it supposed to work? | + | *Observe osc doesn't have slit like spare to keep in place...how is it supposed to work? |
− | ** Is this a DRM thing of sort where they didn't supply the right adapter if the user didn't purchase osc option? | + | **Is this a DRM thing of sort where they didn't supply the right adapter if the user didn't purchase osc option? |
− | ** ie the motor was intentionally disabled and also mechanically hindered | + | **ie the motor was intentionally disabled and also mechanically hindered<br /> |
+ | |||
+ | ===Spare=== | ||
+ | 2019-03-20 | ||
+ | |||
+ | * Try to get old TechPrep base. Negotiations failed | ||
+ | |||
+ | |||
+ | 2019-12-20 | ||
+ | |||
+ | * Won old TechPrep base machine. Change plan from retrofitting SBT machine to actually fitting on this machine | ||
+ | * Without reference machine a little difficult to piece together | ||
− | |||
− | |||
2020-02-23 | 2020-02-23 | ||
− | * Made cable to hook up head | + | *Used reference machine to greatly accelerate repairs |
− | * E1.43 (error code?) persists, but machine seems to work fine | + | *Made cable to hook up head |
+ | *E1.43 (error code?) persists, but machine seems to work fine | ||
<br /> | <br /> |
Revision as of 06:46, 24 February 2020
Motor
Both motors (drive, osc) were identical
PITTMAN GM8712-41 19.1 VDC 187:1 RATIO MFG 05-29-98 S
33 and 17 ohm. Is one of these damaged? Why are their resistances so different?
Tried testing under load
- black goop
- 12.19 V @ 1.266 A (current limited)
- 8 ohm after
- other
- Initially drawing way less current than first
- 12.12 V @ 0.332 A (voltage limited)
- then started smoking
- 11.68 V @ 1.248 A (current limited)
- 6 ohm after
- Initially drawing way less current than first
Aren't these rated 19.1 V? Am I misunderstanding?
Try pinning out the spare bad head
- 1-5: 2.8 k
- 2-6: 147 R
- weird...
what does the secondary machine test? (believe good)
- 1-5: 9.0 R
- 2-6: 10.4 R
What is max voltage allied puts out on motors?
- 1-5
- 7.6 (fixed)
- osc
- 2-6
- 15.0 (max)
- drive
Head connector
Head plug
- 4x2 (8 pins)
- 3 mm pitch
- Head female, cable male
- Maybe Minitek Pwr™ 3.0 Series
- https://www.digikey.com/product-detail/en/amphenol-icc-fci/10132445-08LF/609-5283-ND/7590945
Mini-amphenol | mini-DIN | Signal | Note |
---|---|---|---|
1 | 1 | ||
2 | 2 | ||
3 | 8 | ||
4 | 7 | ||
5 | 3 | ||
6 | 5 | ||
7 | 6 | ||
8 | N/C |
Mini-amphenol | mini-DIN | Signal | Note |
---|---|---|---|
1 | |||
2 | |||
3 | |||
4 | |||
5 | |||
6 | |||
7 | |||
8 |
Motor on mini-DIN
- 1: drive motor +
- 3: drive motor -
- 7 V
Mini-amphenol | cpc | Signal | Note |
---|---|---|---|
1 | 8 | ||
2 | 6 | ||
3 | 1 | ||
4 | 4 | ||
5 | 5 | ||
6 | 3 | ||
7 | 2 | ||
8 | N/C |
McMaster
2020-01-14
- Win cross sectioning kit
2020-02-20
- Order sample holder
- Order replacement tubing
- Later found I had some spare, but meh can use for other projects
Primary
2019-12-19
- Submitted offer. Negotiations failed
2020-02-19
- Seller accepted offer (was firm at old offer). Local pick up
2020-02-232
- Swap DRO from spare to primary machine
2020-02-23
- Observe osc motor is not moving
- Observe osc doesn't have slit like spare to keep in place...how is it supposed to work?
- Is this a DRM thing of sort where they didn't supply the right adapter if the user didn't purchase osc option?
- ie the motor was intentionally disabled and also mechanically hindered
Spare
2019-03-20
- Try to get old TechPrep base. Negotiations failed
2019-12-20
- Won old TechPrep base machine. Change plan from retrofitting SBT machine to actually fitting on this machine
- Without reference machine a little difficult to piece together
2020-02-23
- Used reference machine to greatly accelerate repairs
- Made cable to hook up head
- E1.43 (error code?) persists, but machine seems to work fine