Difference between revisions of "NWR EzLaze"

From Microwiki
Jump to: navigation, search
(mcmaster unit)
(Add Parco and Nikon Engineering to the possible objective list)
 
(32 intermediate revisions by 2 users not shown)
Line 1: Line 1:
 +
John McMaster's notes from repairing and calibrating misc units. For more information contact john@mcmaster.tech
 +
 
Note: New Wave Research was acquired by Electro Scientific Industries, Inc. (ESI)
 
Note: New Wave Research was acquired by Electro Scientific Industries, Inc. (ESI)
  
= ezlaze =
+
==Tilt plate==
 +
These often get separated from units so we made some reproductions. Want one? Contact john@mcmaster.tech
 +
 
 +
 
 +
Variations:
 +
 
 +
*An LCS-II: observe 2 mounting screws
 +
*An ezlaze: observe 3 or so mounting screws
 +
 
 +
 
 +
An LCS-II unit:
 +
 
 +
*Ball bearing
 +
**4.73 mm (0.19" => probably 3/16")
 +
**[https://www.mcmaster.com/9528K13/ Maybe this]
 +
*Pivot set screws
 +
**Qty: 2
 +
**Use 2.5 mm driver
 +
**Special machined screw w/ bearing on end
 +
**Bearing: 3.3 mm OD (0.13")
 +
***[https://www.mcmaster.com/9528K226/ Sub bearing]?
 +
**Thread: M5
 +
**Total length: 9.3 mm
 +
**Threaded length: 7.2 mm
 +
**Material: stainless
 +
**Kevin: glue ball bearing on the end
 +
***4 mm bearings
 +
*Spring SHCS
 +
**Qty: 2
 +
**M4 x 20 mm
 +
**Material: stainless
 +
**Belleville washers: 14
 +
***Arrange as 7 pairs stacked against each other
 +
***[https://www.mcmaster.com/9712K14/ Maybe this] would work?
 +
***Material thickness: 0.46 mm (0.018")
 +
***Total thickness: 0.67 mm (0.026")
 +
***ID: 4.33 mm (0.17")
 +
***OD: 8.4 mm (0.33")
 +
*Set screws
 +
**Qty: 2
 +
**Use 2.0 mm driver
 +
**M4 x 8 mm
 +
**Material: black steel
 +
 
 +
==Objective==
 +
Canonical objective for these is Mitutoyo M Plan Apo. Could something else work?
 +
 
 +
 
 +
Candidates:
 +
 
 +
*Search terms
 +
**"M Plan Apo"
 +
**LWD objective
 +
**long working distance objective
 +
*Mitutoyo M Plan Apo
 +
*B&L Industrial
 +
**Wrong thread
 +
**Not infinity
 +
**Seem to have a weird design, as explained in the [https://worldwide.espacenet.com/patent/search/family/024876339/publication/US3471218A?q=pn%3DUS3471218A US3471218A] patent
 +
**...but cheap and might be good enough
 +
*Optem
 +
*Seiwa
 +
*Sigma Koki
 +
*USMCO
 +
*USMC
 +
*V-Technology
 +
*Parco
 +
*Nikon Engineering
 +
*New Wave
 +
**As in NWR? Are these intended for ezlaze?
 +
*Correct Tokyo (Seiwa)
 +
*Olympus BHM
 +
**Correct thread size
 +
**Very cheap
 +
**Incorrect focus length (180mm, instead of 200mm)
 +
**LWD variants
 +
***LWD M S Plan
 +
****20x (WD=7mm)
 +
****50x (WD=3mm)
 +
**non-LWD variants with maybe enough working distance (only 20x and up)
 +
***M S Plan
 +
****20x (WD=3mm)
 +
****50x (WD=0.47mm)
 +
***M D Plan
 +
****20x (WD=1mm)
 +
***M D Ach
 +
****20x (WD=1.1mm)
 +
****50x (WD=0.45mm)
 +
***Neo S Plan
 +
****20x (WD=3mm)
 +
****50x (WD=0.5mm)
 +
***Neo D Plan
 +
****20x (WD=1mm)
 +
 
 +
 
 +
To test:
 +
 
 +
*Inspect objective
 +
*Run image quality test (MTF?)
 +
*Long duration test...maybe 1000 shots at max power?
 +
*Inspect objective
 +
*Run image quality test (MTF?)
 +
 
 +
<br />
 +
 
 +
===Results===
 +
 
 +
 
 +
Compatible:
 +
 
 +
*Mitutoyo M Plan Apo
 +
 
 +
 
 +
Okay (seem compatible from initial testing, but no extensive tests done):
 +
 
 +
*...
 +
 
 +
 
 +
Incompatible:
 +
 
 +
*Edmund Optics HR (source: McMaster tried the 10x)
 +
 
 +
<br />
 +
==Cabling==
 +
DB25
 +
 
 +
*Has a slot instead of a round cable hole
 +
*3M 3485-2300
 +
 
 +
 
 +
Power connector
 +
 
 +
*Got blank shell and cut slot w/ hacksaw. Then melted new plastic in to close up the side
 +
 
 +
==Noisy/slow crystal/filter selection investigation==
 +
Symptom: the filter wheel that contains wavelength selection and high/low takes longer to settle than usual and overshoots a few times before settling down
 +
 
 +
Thoughts:
 +
 
 +
*Motor problem?
 +
*Pot problem?
 +
*PID parameters can be adjusted? Are there any trim pots?
 +
*No spare units to scavenge parts off of
 +
 
 +
Motor
 +
 
 +
*1616E012S T 265
 +
 
 +
==Repair / calibration==
 +
 
 +
===Uneven cut===
 +
 
 +
  With the uneven cut, is possible the laser is misalign and will need to
 +
  be optimize and realign.  The possible misalignment can cause by the
 +
  Q-SW (pocket cell) is deteriorating and may need replace.  For the XY
 +
  aperture shutter, we will replace both Motors and Potentiometer.
 +
 
 +
===Flashlamp replace===
 +
 
 +
Flashlamp rated for 30 million shots. Skeptical replacing it was going to help, but tried anyway. Didn't seem to improve power output. Maybe its the uneven cut issue above?
 +
 
 +
7/32" allen key for side of flashlamp housing
 +
 
 +
2 mm for screwing cover in for the long buried screw near the HV generator
 +
 
 +
1.3 mm allen for the short ones
 +
 
 +
mcmaster original lamp says: LL7039/88
 +
 
 +
Lamp part numbers:
 +
 
 +
*New Wave Research 0002-0036
 +
*LL7039
 +
**Printed on lamp: LL7039/88
 +
**Who's part number is this?
 +
**Total length: ~78 mm
 +
**Arc length: ~46 mm
 +
**Diameter:  ~4.97 mm
 +
 
 +
{| class="wikitable"
 +
|+
 +
!Vendor / part
 +
!Date
 +
!Price
 +
!Lead time
 +
|-
 +
|1 EU, opt 1
 +
|2019-09
 +
|£122.00
 +
|10 weeks
 +
|-
 +
|1 EU, opt 2
 +
|2019-09
 +
|£185.70
 +
|10-12 weeks
 +
|-
 +
|1 US
 +
|2009-12
 +
|$215
 +
|2 days
 +
May be wrong part. Not sure if billing is correct
 +
|-
 +
|2
 +
|2019-09
 +
|$245
 +
|2 days
 +
|-
 +
|3
 +
|2019-09
 +
|$197.36
 +
|2 days
 +
|}
 +
Spare lamp box says "Lamp no: LL7039"
 +
 
 +
*https://usermanual.wiki/Document/LaserLampsIPLlampls.1504005633/view: Laser SOS Group, Continuum Minilite 10, 20, early models, 203–0037
 +
 
 +
Suggested vendors:
 +
 
 +
*https://www.tjslasers.com/flash-lamp-flashlamps-Nd-YAG-Lasers
 +
*https://www.lasersos.com/wp-content/uploads/2018/04/Laser-Lamps-catalog.pdf
 +
**LL6811 Medilite 1
 +
**LL7886 Solo PIV 0002-0036, Tempest, Gemini
 +
 
 +
===2019-09-25===
 +
Visit Kevin, got:
 +
 
 +
*2x ezlaze laser head
 +
*1x ESI power meter head
 +
*1x power meter head w/ controller
 +
 
 +
 
 +
Tests on my ezlaze:
 +
 
 +
*DB25 only: motors twitch at startup. Interlock lit red, cannot turn on
 +
*Both connectors: motors twitch at startup. Interlock lit red. However, press "ON" moves from red interlock to yellow emission
 +
*When hitting "ON", a loud relay is heard in the controller
 +
 
 +
Kevin-1 (K1):
 +
 
 +
*Missing tilt plate
 +
*Missing tilt plate screws
 +
*No camera adapter
 +
*Motors do not twitch at power up
 +
*Unable to bring out of interlock
 +
*Missing shutter
 +
*LCSII....ah not exactly the same
 +
 
 +
Kevin-2 (K2)
  
== mcmaster unit ==
+
*Motors twitch at startup
 +
*Bent power connector. Was able to bend back into shape
 +
*Mounting plate screws but no mounting plate
 +
*Has shutter
 +
*EZLAZE 532/355
 +
*Beam profile uneven (as seen through laser glasses). TODO: point at camera?
 +
*Sensor 3" from laser, 10 sec average
 +
**500L: 65 uJ
 +
**1000L: 124 uJ
 +
**500H: 192 uJ
 +
**1000H: 382 uJ
 +
 
 +
 
 +
ESI power meter
 +
 
 +
*Battery: Everyday 504, NEDA 2220, 15V
 +
 
 +
===2019-10-28===
 +
Documenting backlog. Especially around Pockells cell / EOM repair
 +
 
 +
Open question about q-switch polarity. Tried putting my (mostly bad) EOM in my laser forwards/backwards and positive/negative polarity. Only forward positive configuration fires. Not sure if polarity reversal is bad
 +
 
 +
Scrapped ezlaze Kevin had in dire shape to get sacrificial q-switch. Housing very different, but core looked the same. Unfortunately, does not seem to work in my original laser. Also tried reversing it, but didn't help. NOTE: Kevin has a second identical loose EOM assembly. We should try it
 +
 
 +
Played briefly with transferring fluid from sacrificial larger EOM. Had issue with bubble, also still unclear if fluid matched correctly
 +
 
 +
 
 +
Pockels cell research:
 +
 
 +
*Probably made by Qioptiq, but may be branded Gsanger or LINOS
 +
 
 +
<blockquote></blockquote><blockquote>Due to a series of acquisitions, Qioptiq</blockquote><blockquote>has an impressive history and pedigree, benefi ting</blockquote><blockquote>from the knowledge and experience of LINOS,</blockquote><blockquote>Point Source, Rodenstock Precision Optics, Spindler</blockquote><blockquote>& Hoyer, Gsänger, Optem, Pilkington, Avimo and</blockquote><blockquote>others.</blockquote>
 +
 
 +
 
 +
Did a lot of research on the fluid. Index matching fluid (IMF) is required due to the high efficiencies required to make a cavity resonate.
 +
 
 +
https://microwiki.org/media/qti/qti_an_dry_pockels_cell_vs_wet_pockels_cells.pdf
 +
 
 +
*Great paper on IMF choices
 +
*TLDR: FC-43 for most applications, but more exotic choices if you don't care about carcinogens and/or cost
 +
 
 +
 
 +
Ordering:
 +
 
 +
*quantumtech.com: "We have index matching fluid $175 for 30ml bottle and available from stock.  We charge for repair would be approximately $400. "
 +
*Tried to order FC-43 from Sigma, but they denied my order
 +
**<nowiki>https://www.sigmaaldrich.com/catalog/product/sial/77299</nowiki>
 +
**Heptacosafluorotributylamine (Fluorinert™ FC-43)
 +
**5 mL: 52.60
 +
**25 mL: 235.00
 +
 
 +
=ezlaze=
 +
 
 +
 
 +
==mcmaster unit==
  
 
[[User:Mcmaster/Brainscope]]
 
[[User:Mcmaster/Brainscope]]
  
== IOA unit ==
+
==IOA unit==
  
 
TODO: add pictures (if someone gives them to me...)
 
TODO: add pictures (if someone gives them to me...)
  
== eBay ntxsupply unit (EZ-LAZE 532/355) ==
+
==eBay ntxsupply unit (EZ-LAZE 532/355)==
  
 
"New Wave Research EZ-LAZE 532 / 355 Power Supply & Control Panel & Foot Pedal"
 
"New Wave Research EZ-LAZE 532 / 355 Power Supply & Control Panel & Foot Pedal"
Line 34: Line 337:
  
  
== eBay outback6 unit (EZ-LAZE 532) ==
+
==eBay outback6 unit (EZ-LAZE 532)==
  
 
"G102838 New Wave Research EZ-LAZE/532 Laser w/Power Supply & Controller"
 
"G102838 New Wave Research EZ-LAZE/532 Laser w/Power Supply & Controller"
Line 66: Line 369:
  
  
== eBay  express_test unit (EZ-LAZE LCS-II 532/355) ==
+
==eBay  express_test unit (EZ-LAZE LCS-II 532/355)==
  
 
[[File:ebay_express_test_400896932451_0.jpg|300 px]]
 
[[File:ebay_express_test_400896932451_0.jpg|300 px]]
Line 73: Line 376:
  
  
= ezlaze2 =
+
=ezlaze2=
  
== cl.cam.ac.uk ==
+
==cl.cam.ac.uk==
  
 
http://www.cl.cam.ac.uk/~sps32/laser_cutter.pdf
 
http://www.cl.cam.ac.uk/~sps32/laser_cutter.pdf
Line 81: Line 384:
 
Has a great teardown and some really useful info about system problems
 
Has a great teardown and some really useful info about system problems
  
= ezlaze3 =
+
=ezlaze3=
  
 
[http://www.eolite.com/Products/SolidStateLasers/EZLazeSeries.aspx#Specifications Source]:
 
[http://www.eolite.com/Products/SolidStateLasers/EZLazeSeries.aspx#Specifications Source]:
Line 87: Line 390:
 
Main Features
 
Main Features
 
     Air-cooled architecture
 
     Air-cooled architecture
## No cooling water
+
 
## Compact, Lightweight
+
##No cooling water
## No maintenance required
+
##Compact, Lightweight
 +
##No maintenance required
 +
 
 
     Selectable wavelengths allow cutting
 
     Selectable wavelengths allow cutting
 
     and machining of a wide range of materials:
 
     and machining of a wide range of materials:
## 1064 nm, 532 nm, 355 nm or 266 nm
+
 
 +
##1064 nm, 532 nm, 355 nm or 266 nm
 +
 
 
     3-axis control of cut size and orientation.
 
     3-axis control of cut size and orientation.
 
     From 2μm x 2μm (with 100X objective)
 
     From 2μm x 2μm (with 100X objective)
Line 105: Line 412:
 
     of FA microscopes (Mitutoyo, Motic, Seiwa).
 
     of FA microscopes (Mitutoyo, Motic, Seiwa).
 
     3 Operating modes:
 
     3 Operating modes:
## Single shot
+
 
## Continuous (1 Hz)
+
##Single shot
## Burst (max. of 10 sec. at 5 Hz).
+
##Continuous (1 Hz)
 +
##Burst (max. of 10 sec. at 5 Hz).
 
 
 
Standard configuration includes:
 
Standard configuration includes:
Line 129: Line 437:
 
     Internal Spot Marker
 
     Internal Spot Marker
 
     Microscope & Accessories
 
     Microscope & Accessories
 +
 +
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 +
<br />
 
</code>
 
</code>
  
  
= QuikLaze =
+
=QuikLaze=
  
 
I asked for a ezlaze manual and was given a quicklaze manual...w/e.  Looks like they share a lot of common components
 
I asked for a ezlaze manual and was given a quicklaze manual...w/e.  Looks like they share a lot of common components
Line 146: Line 472:
 
The quicklaze power supply looks pretty similar to the ezlaze power supply...possibly compatibile
 
The quicklaze power supply looks pretty similar to the ezlaze power supply...possibly compatibile
  
== eBay surplusexiting complete 532 ==
+
==eBay surplusexiting complete 532==
  
 
Overview:
 
Overview:
Line 171: Line 497:
  
  
== eBay surplusexiting prism_electronics2 bad power supply ==
+
==eBay surplusexiting prism_electronics2 bad power supply==
  
 
[[File:ebay_prism_electronics2_291003612067_0.jpg|300 px]]
 
[[File:ebay_prism_electronics2_291003612067_0.jpg|300 px]]
Line 180: Line 506:
 
[[File:ebay_prism_electronics2_291003612067_5.jpg|300 px]]
 
[[File:ebay_prism_electronics2_291003612067_5.jpg|300 px]]
  
== eBay s.scherets ==
+
==eBay s.scherets==
  
 
A less common stand alone system.  Rackmount power supply
 
A less common stand alone system.  Rackmount power supply
Line 189: Line 515:
 
[[File:ebay_s.scherets_171042385856_1.jpg|300 px]]
 
[[File:ebay_s.scherets_171042385856_1.jpg|300 px]]
  
= References =
+
=References=
  
* http://www.esi.com/Products/Semiconductor/FlatPanelDisplayRepairFailureAnalysis/EzLaze.aspx
+
*http://www.esi.com/Products/Semiconductor/FlatPanelDisplayRepairFailureAnalysis/EzLaze.aspx
* http://www.eolite.com/Products/SolidStateLasers/EZLazeSeries.aspx#Overview
+
*http://www.eolite.com/Products/SolidStateLasers/EZLazeSeries.aspx#Overview
* https://www.youtube.com/watch?v=RhLxaHyNIv4
+
*https://www.youtube.com/watch?v=RhLxaHyNIv4
* Internals working
+
**Internals working
 +
*https://microwiki.org/media/nwr/
 +
*https://microwiki.org/media/continuum/

Latest revision as of 23:56, 15 January 2023

John McMaster's notes from repairing and calibrating misc units. For more information contact john@mcmaster.tech

Note: New Wave Research was acquired by Electro Scientific Industries, Inc. (ESI)

Tilt plate

These often get separated from units so we made some reproductions. Want one? Contact john@mcmaster.tech


Variations:

  • An LCS-II: observe 2 mounting screws
  • An ezlaze: observe 3 or so mounting screws


An LCS-II unit:

  • Ball bearing
  • Pivot set screws
    • Qty: 2
    • Use 2.5 mm driver
    • Special machined screw w/ bearing on end
    • Bearing: 3.3 mm OD (0.13")
    • Thread: M5
    • Total length: 9.3 mm
    • Threaded length: 7.2 mm
    • Material: stainless
    • Kevin: glue ball bearing on the end
      • 4 mm bearings
  • Spring SHCS
    • Qty: 2
    • M4 x 20 mm
    • Material: stainless
    • Belleville washers: 14
      • Arrange as 7 pairs stacked against each other
      • Maybe this would work?
      • Material thickness: 0.46 mm (0.018")
      • Total thickness: 0.67 mm (0.026")
      • ID: 4.33 mm (0.17")
      • OD: 8.4 mm (0.33")
  • Set screws
    • Qty: 2
    • Use 2.0 mm driver
    • M4 x 8 mm
    • Material: black steel

Objective

Canonical objective for these is Mitutoyo M Plan Apo. Could something else work?


Candidates:

  • Search terms
    • "M Plan Apo"
    • LWD objective
    • long working distance objective
  • Mitutoyo M Plan Apo
  • B&L Industrial
    • Wrong thread
    • Not infinity
    • Seem to have a weird design, as explained in the US3471218A patent
    • ...but cheap and might be good enough
  • Optem
  • Seiwa
  • Sigma Koki
  • USMCO
  • USMC
  • V-Technology
  • Parco
  • Nikon Engineering
  • New Wave
    • As in NWR? Are these intended for ezlaze?
  • Correct Tokyo (Seiwa)
  • Olympus BHM
    • Correct thread size
    • Very cheap
    • Incorrect focus length (180mm, instead of 200mm)
    • LWD variants
      • LWD M S Plan
        • 20x (WD=7mm)
        • 50x (WD=3mm)
    • non-LWD variants with maybe enough working distance (only 20x and up)
      • M S Plan
        • 20x (WD=3mm)
        • 50x (WD=0.47mm)
      • M D Plan
        • 20x (WD=1mm)
      • M D Ach
        • 20x (WD=1.1mm)
        • 50x (WD=0.45mm)
      • Neo S Plan
        • 20x (WD=3mm)
        • 50x (WD=0.5mm)
      • Neo D Plan
        • 20x (WD=1mm)


To test:

  • Inspect objective
  • Run image quality test (MTF?)
  • Long duration test...maybe 1000 shots at max power?
  • Inspect objective
  • Run image quality test (MTF?)


Results

Compatible:

  • Mitutoyo M Plan Apo


Okay (seem compatible from initial testing, but no extensive tests done):

  • ...


Incompatible:

  • Edmund Optics HR (source: McMaster tried the 10x)


Cabling

DB25

  • Has a slot instead of a round cable hole
  • 3M 3485-2300


Power connector

  • Got blank shell and cut slot w/ hacksaw. Then melted new plastic in to close up the side

Noisy/slow crystal/filter selection investigation

Symptom: the filter wheel that contains wavelength selection and high/low takes longer to settle than usual and overshoots a few times before settling down

Thoughts:

  • Motor problem?
  • Pot problem?
  • PID parameters can be adjusted? Are there any trim pots?
  • No spare units to scavenge parts off of

Motor

  • 1616E012S T 265

Repair / calibration

Uneven cut

 With the uneven cut, is possible the laser is misalign and will need to
 be optimize and realign.  The possible misalignment can cause by the
 Q-SW (pocket cell) is deteriorating and may need replace.  For the XY
 aperture shutter, we will replace both Motors and Potentiometer.

Flashlamp replace

Flashlamp rated for 30 million shots. Skeptical replacing it was going to help, but tried anyway. Didn't seem to improve power output. Maybe its the uneven cut issue above?

7/32" allen key for side of flashlamp housing

2 mm for screwing cover in for the long buried screw near the HV generator

1.3 mm allen for the short ones

mcmaster original lamp says: LL7039/88

Lamp part numbers:

  • New Wave Research 0002-0036
  • LL7039
    • Printed on lamp: LL7039/88
    • Who's part number is this?
    • Total length: ~78 mm
    • Arc length: ~46 mm
    • Diameter: ~4.97 mm
Vendor / part Date Price Lead time
1 EU, opt 1 2019-09 £122.00 10 weeks
1 EU, opt 2 2019-09 £185.70 10-12 weeks
1 US 2009-12 $215 2 days

May be wrong part. Not sure if billing is correct

2 2019-09 $245 2 days
3 2019-09 $197.36 2 days

Spare lamp box says "Lamp no: LL7039"

Suggested vendors:

2019-09-25

Visit Kevin, got:

  • 2x ezlaze laser head
  • 1x ESI power meter head
  • 1x power meter head w/ controller


Tests on my ezlaze:

  • DB25 only: motors twitch at startup. Interlock lit red, cannot turn on
  • Both connectors: motors twitch at startup. Interlock lit red. However, press "ON" moves from red interlock to yellow emission
  • When hitting "ON", a loud relay is heard in the controller

Kevin-1 (K1):

  • Missing tilt plate
  • Missing tilt plate screws
  • No camera adapter
  • Motors do not twitch at power up
  • Unable to bring out of interlock
  • Missing shutter
  • LCSII....ah not exactly the same

Kevin-2 (K2)

  • Motors twitch at startup
  • Bent power connector. Was able to bend back into shape
  • Mounting plate screws but no mounting plate
  • Has shutter
  • EZLAZE 532/355
  • Beam profile uneven (as seen through laser glasses). TODO: point at camera?
  • Sensor 3" from laser, 10 sec average
    • 500L: 65 uJ
    • 1000L: 124 uJ
    • 500H: 192 uJ
    • 1000H: 382 uJ


ESI power meter

  • Battery: Everyday 504, NEDA 2220, 15V

2019-10-28

Documenting backlog. Especially around Pockells cell / EOM repair

Open question about q-switch polarity. Tried putting my (mostly bad) EOM in my laser forwards/backwards and positive/negative polarity. Only forward positive configuration fires. Not sure if polarity reversal is bad

Scrapped ezlaze Kevin had in dire shape to get sacrificial q-switch. Housing very different, but core looked the same. Unfortunately, does not seem to work in my original laser. Also tried reversing it, but didn't help. NOTE: Kevin has a second identical loose EOM assembly. We should try it

Played briefly with transferring fluid from sacrificial larger EOM. Had issue with bubble, also still unclear if fluid matched correctly


Pockels cell research:

  • Probably made by Qioptiq, but may be branded Gsanger or LINOS

Due to a series of acquisitions, Qioptiq

has an impressive history and pedigree, benefi ting

from the knowledge and experience of LINOS,

Point Source, Rodenstock Precision Optics, Spindler

& Hoyer, Gsänger, Optem, Pilkington, Avimo and

others.


Did a lot of research on the fluid. Index matching fluid (IMF) is required due to the high efficiencies required to make a cavity resonate.

https://microwiki.org/media/qti/qti_an_dry_pockels_cell_vs_wet_pockels_cells.pdf

  • Great paper on IMF choices
  • TLDR: FC-43 for most applications, but more exotic choices if you don't care about carcinogens and/or cost


Ordering:

  • quantumtech.com: "We have index matching fluid $175 for 30ml bottle and available from stock. We charge for repair would be approximately $400. "
  • Tried to order FC-43 from Sigma, but they denied my order
    • https://www.sigmaaldrich.com/catalog/product/sial/77299
    • Heptacosafluorotributylamine (Fluorinert™ FC-43)
    • 5 mL: 52.60
    • 25 mL: 235.00

ezlaze

mcmaster unit

User:Mcmaster/Brainscope

IOA unit

TODO: add pictures (if someone gives them to me...)

eBay ntxsupply unit (EZ-LAZE 532/355)

"New Wave Research EZ-LAZE 532 / 355 Power Supply & Control Panel & Foot Pedal"

Need power supply, this unit sounds very similar to mine and has the right connections. In negotiations to try to get this.

Overview:

Ebay ntxsupply 151675355993 0.jpg

Control box:

Ebay ntxsupply 151675355993 1.jpg Ebay ntxsupply 151675355993 2.jpg Ebay ntxsupply 151675355993 3.jpg

Power supply:

Ebay ntxsupply 151675355993 4.jpg Ebay ntxsupply 151675355993 5.jpg Ebay ntxsupply 151675355993 6.jpg


eBay outback6 unit (EZ-LAZE 532)

"G102838 New Wave Research EZ-LAZE/532 Laser w/Power Supply & Controller"

A possibility, but its a complete system which means it will command a higher price...probably harder to break it up just to get the PS

Overview:

Ebay outback6 331147302083 0.jpg

Controller:

Ebay outback6 331147302083 1.jpg

Pedal:

Ebay outback6 331147302083 3.jpg

Power supply:

Ebay outback6 331147302083 9.jpg Ebay outback6 331147302083 2.jpg

Laser head:

Ebay outback6 331147302083 4.jpg Ebay outback6 331147302083 5.jpg Ebay outback6 331147302083 6.jpg Ebay outback6 331147302083 7.jpg Ebay outback6 331147302083 8.jpg


eBay express_test unit (EZ-LAZE LCS-II 532/355)

Ebay express test 400896932451 0.jpg

System looks a little different


ezlaze2

cl.cam.ac.uk

http://www.cl.cam.ac.uk/~sps32/laser_cutter.pdf

Has a great teardown and some really useful info about system problems

ezlaze3

Source: Main Features

   Air-cooled architecture
    1. No cooling water
    2. Compact, Lightweight
    3. No maintenance required
   Selectable wavelengths allow cutting
   and machining of a wide range of materials:
    1. 1064 nm, 532 nm, 355 nm or 266 nm
   3-axis control of cut size and orientation.
   From 2μm x 2μm (with 100X objective)
   to 50μm x 50μm (with 50X objective).
   Rotation through 180°.
   Simple operation via remote control box
   with a 6.5cm x 7.5cm LCD screen displaying
   menu-based commands or via RS232 interface.
   Precise energy control over a wide range.
   Internal LED spot marker.
   Compatible with most major brands
   of FA microscopes (Mitutoyo, Motic, Seiwa).
   3 Operating modes:
    1. Single shot
    2. Continuous (1 Hz)
    3. Burst (max. of 10 sec. at 5 Hz).

Standard configuration includes:

   Laser Head
   Standard Energy
   XY Shutter (Std. Resolution)
   Power Supply
   Remote Control
   2.4m umbilical
   Video spot marker
   Foot switch

Options:

   High Pulse Energy
   Rotating Shutter
   High Resolution Shutter
   Motorized Polarizer
   Horizontal Mount Laser
   Head with Robot-length
   Umbilical
   Internal Spot Marker
   Microscope & Accessories



















QuikLaze

I asked for a ezlaze manual and was given a quicklaze manual...w/e. Looks like they share a lot of common components

Interesting excerpt from manual: The QuikLaze is designed to operate on specially modified Mitutoyo FS60Y and FS50 and Ready products A-Zoom and S-Scope microscopes. Become familiar with your microscope before attempting to install the QuikLaze.

The quicklaze power supply looks pretty similar to the ezlaze power supply...possibly compatibile

eBay surplusexiting complete 532

Overview:

Ebay surplusexiting 291461826013 0.jpg Ebay surplusexiting 291461826013 9.jpg Ebay surplusexiting 291461826013 4.jpg

Power supply:

Ebay surplusexiting 291461826013 1.jpg Ebay surplusexiting 291461826013 5.jpg

Controller:

Ebay surplusexiting 291461826013 2.jpg Ebay surplusexiting 291461826013 3.jpg

Head:

Ebay surplusexiting 291461826013 6.jpg Ebay surplusexiting 291461826013 7.jpg Ebay surplusexiting 291461826013 8.jpg


eBay surplusexiting prism_electronics2 bad power supply

Ebay prism electronics2 291003612067 0.jpg Ebay prism electronics2 291003612067 1.jpg Ebay prism electronics2 291003612067 2.jpg Ebay prism electronics2 291003612067 3.jpg Ebay prism electronics2 291003612067 4.jpg Ebay prism electronics2 291003612067 5.jpg

eBay s.scherets

A less common stand alone system. Rackmount power supply

They called it a quicklaze but I think its actually an ezlaze?

Ebay s.scherets 171042385856 0.jpg Ebay s.scherets 171042385856 1.jpg

References